본문바로가기

Medical

We will do our best for customer's satisfaction

Small Steam Sterilizer

  • MEDICAL
  • Small Steam Sterilizer
  • HS-85SL

HS-85SL

Vertical type, Automatic Door, LCD Monitor type

  • close
  • open
  • open_moving

  • close
  • open
  • open_moving

Highlights
Scroll down

  • 01 Far more convenient and safe sterilization by motor driven Automatic Door

    This sterilizer is adopted a unique vertical type cylindrical chamber equipped with a motor driven automatic door, therefore the chamber loading, unloading and sterilizer operation are far more convenient and safe compare to the manual door. The vertical type chamber allows to use a wider chamber than another small sterilizers in the confined space, and is the most effective for sterilizing a liquid load, more useful for liquid contained in a big glass bottle at the clinical laboratories and research institutes. The operator could check the cycle progressing conveniently on the LCD monitor because the state of sterilizer operation is displayed digitally in real time on it.

  • 02 Various sterilization cycle and self-diagnosis program

    5 kinds of basic cycle program for sterilizing in accordance with items to be sterilized and packaging type, therefore the operator could select a proper cycle program and perform sterilization conveniently. Especially the liquid contained in the glass bottles could be sterilized safely and efficiently, and users can change the parameters of basic cycle programs at user's pleasure for handling a particular sterilizing condition. If an error has occurred during operating the sterilizer, the error code is displayed on the LCD monitor accompanying an audible alarm and the cycle in progress is aborted at the same time automatically for the safety of workplace by the self-diagnosis program.

  • 03 Cycle Execution Data Storing

    The cycle performance records executed at least for 11 years are stored in the control unit in order. The user can automatically search the results of sterilization cycle performed previously by input the date performed cycle or the cycle count number and prints out the result record by the built-in panel printer.

  • 04 Unique exterior small steam generator

    The built-in exterior small steam generator which is newly-developed by our advanced technology can vaporize feed water supplied from the reservoir instantly and the generated steam is injected into the chamber as a sterilizing media, therefore the chamber is less contaminated than the internal steam generating type and the usable loading space is widened because there is no heater assembly in the chamber.

  • 05 Use of clean steam every cycle

    The used water after completion of a sterilizing cycle is collected in the collection container designed separately with the reservoir, therefore the saturated steam generated by a clean water always can be supplied to perform every sterilizing cycle.

  • 06 Safety Equipments

    This sterilizer has the safety system for the safety of workplace designed that there exists more pressure compare to the set level in the chamber by the pressure switch. The safety valve operates and releases chamber pressure automatically when an overpressure more than the limited value has been formed in the chamber for a certain reason.

  • 07 Panel Printer

    The built-in panel printer prints out the cycle performance record for securing the traceability afterwards.

Specifications

Model HS-85SL
Overall Dimension W600 x D705 x H992 mm
Chamber Vertical type, IDø400 x L741 mm, Vol: 85ℓ, Material: STS 316L
Max. Working Pressure : 2.40bar, Min. Working Pressure : -1.00bar
Max. Working Temperature : 138.0℃, Test Pressure : 5.13bar

Loading Device

Cylindrical Stainless Steel Basket
ID∅355 x L215 mm 3ea 

Reservoir W258 x D130 x H256 mm, Vol: 7.5ℓ (full water level)
Collection Container Vol: 5.0ℓ (full water level)
Operating Temp. Ranger 121.0℃ ~ 135.0℃
Sterilizing Agent

Saturated Steam

Air Removal Method Vacuum
Retort Sterilizers R Type Sterilizers (optional)
Air Removal Equipment Vacuum Pump (Diaphragm type)

Air Filter

Filtration efficiency: 99.997% for 0.3㎛ particles
Door Operation Automatic Driving by Motor
Controls Microprocessor
Display Graphic LCD(LED back-lighting)

Panel Printer

Direct Thermal Printing, Printing paper roll; W58 x ∅40 mm, L=13 m
Power Source AC 230V, 50/60Hz
Power Consumption 3,300 W
Option Water Saving Device / Cooling Device
Net Weight 140 kg
Environment Conditions ㆍ Altitudes up to 2,000 meters
ㆍ Temperature range of +5℃ to +40℃
ㆍ Maximum relative humidity 80% for temperature up to 31℃ decreasing linearly to 50% relative humidity at 40℃
ㆍ Main supply voltage fluctuation of ±10% of nominal
ㆍ Installation Category (Overvoltage Category) Ⅱ, Pollution Degree 2