Large Steam Sterilizer

  • MEDICAL
  • Large Steam Sterilizer
  • HS-1000R

HS-1000R

  • The front (close)
  • The rear (close)
  • The front (open)
  • Double-Door-,-open

  • The front (close)
  • The rear (close)
  • The front (open)
  • Double-Door-,-open

Highlights
Scroll down

  • 01

    High-pressure saturated steam supplied from the outside is used as a sterilization medium, and various retort pouches and blister package products are safely sterilized without damage through a cycle program dedicated to retort. In addition, 5 types of basic cycle programs that can sterilize general medical equipment are built-in, and the user can arbitrarily set and sterilize the program according to various situations, so the equipment's effectiveness is high.

  • 02

    The sterilization process is fully automatically controlled by the microprocessor, and the sterilizer can be easily operated and controlled while viewing the progress of the sterilization cycle directly on the touch screen type color LCD monitor. The built-in self-diagnosis program monitors the operation status of the sterilizer in real time and automatically takes safety measures such as sound alarm, cycle stop, and error message display when an abnormality occurs during operation, so that operation errors or malfunctions can be dealt with promptly.

  • 03

    The touch screen type color LCD allows the user to operate the sterilizer easily by light touching the screen buttons and realizes delicate graphic display.

  • 04

    As a unique feature of our sterilizer, at least 11 years of cycle execution records are automatically stored in the Flash memory of controller. The user can automatically retrieve the result of the sterilization cycle that was executed in the past as needed and print the contents with the panel printer mounted on the sterilizer, and use the result as an evidence of sterilization quality.

  • 05

    By printing cycle execution data such as temperature and pressure, sterilization time, drying time, and error by hour in the sterilization process with a panel printer, it can be stored and referred to as sterilization quality data.

  • 06

    The temperature and pressure are controlled automatically to be suitable for the sterilization condition, their changing values through cycle progressing are displayed on the screen in real time by graph, and saved in the memory device.

  • 07

    The operating status is diagnosed, displayed and stored in real time, so that an operation error or a trouble can be recovered rapidly.

  • 08

    The standard language displayed on the screen is Korean language and English (another language can be provided as an optional specification).

Specifications

Model HS-1000R
Style

Single Door

Double Door

Overall Dimension W1,385 x D1,705 x H1,894 ㎜ W1,385 x D2,036 x H1,894 ㎜
Chamber W680 x D1,286 x H1,200 ㎜
Vol: 1,049ℓ, Material: STS 316L
W680 x D1,316 x H1,200 ㎜
Vol: 1,073ℓ, Material: STS 316L
Door Operation Power(Automatic) Operation/Manual Operation
Loading System Loading Car: W600 x D1,200 x H1,072㎜
Transfer Carriage: W560 x D1,490 x H986㎜
Operating Temperature 121.0 ℃ ~ 135.0 ℃
Air Removal Vacuum / Gravity
Controls Microprocessor
Display Touch Screen type Color LCD
Panel Printer Direct Thermal Printing, Printing paper roll: W58 x Ø40 mm(length:13mm)
Power Source AC 230V, 1ph, 50/60Hz, 15A
Steam Source Saturated Steam (Dry Rate: more than 97%) , Pressure: 3.50~5.00bar,
Connection Pipe: Unit Side 1″ FPT, Facility Side 1-1/4" FPT
Water Source Soft Water less than 20℃, Pressure: 2.10~5.00bar,
Connection Pipe: Unit Side 1" FPT, Facility Side 1-1/4" FPT
Vacuum System Water Ejector, Water Pump
Drainage Connection Pipe : Unit Side 1-1/4″ FPT
Exclusive Drainage : 2″ MPT(Metallic Pipe)
Net Weight Unit : 1,656㎏
Loading Car: 58㎏ x 1
Carriage: 52㎏ x 1
Unit: 2,279㎏
Loading Car: 58㎏ x 1
Carriage: 52㎏ x 2
Environment Conditions(IEC61010-1) ㆍ Attude up to 2,000m
ㆍ Temperature range of +5°C to +40°C
ㆍ Main supply voltage fluctuation of ±10% of nominal
ㆍ Installation Category (Overvoltage Category) II, Polution Degee: 2
ㆍ Maximum relative humidity of 80% for temperature up to 31°C decreasing linearly to 50% relative humidity at 40°C