E.O. Gas Sterilizer

  • MEDICAL
  • E.O. Gas Sterilizer
  • HG-1000E·P

HG-1000E·P

  • Close
  • Open

  • Close
  • Open

Highlights
Scroll down

  • 01

    This 100% ethylene oxide(EO) gas sterilizer is intended to sterilize the medical instruments, devices and supplies, which are labile to high temperature and
    pressure and improper for steam sterilization structurally, safely and effectively at
    low temperature.

  • 02

    EO gas sterilization is complementary to the steam sterilization and is the essential sterilizing method for the hospitals and the medical institutions employing
    precise and delicate medical instruments and devices.

  • 03

    This sterilizer uses the "100% ethylene oxide(EO) gas" as sterilant and can sterilize the medical devices, instruments and supplies effectively by convenient
    way of only inserting a small disposable gas cartridge on the holder recessed in
    the chamber wall.

  • 04

    Especially the sterilization and the aeration process are all completed in the chamber under negative pressure, therefore leakage of EO gas in the working area
    can be fully prevented. EO gas being sucked by powerful vacuum pump is
    exhausted to the drain system with water safely after sterilization phase.

  • 05

    4 kinds of sterilization cycle program and the vacuum leak test cycle program are embedded in the sterilizer and the operator can select a proper cycle program
    according to the type of materials or items to be sterilized and sterilize them
    conveniently. User can set a special user cycle program by changing the basic cycle parameters at user's pleasure for handling a particular sterilization condition
    smoothly if needed.

  • 06

    The self diagnostic program monitors the cycle progress in real time, when an error has occurred during progressing a cycle, displays the corresponding error
    code and message on the touch screen immediately accompanying an audible
    alarm, and aborts the cycle in progress at the same time after venting the
    chamber automatically for safety of workplace.

  • 07

    At least 11 years of cycle execution records are automatically stored in the Flash memory of controller. The user can automatically retrieve the result of the
    sterilization cycle that was executed in the past as needed and print the contents
    with the panel printer mounted on the sterilizer, and use the result as an evidence
    of sterilization quality.

  • 08

    Touch screen type color LCD allows user to operate the sterilizer more easily and quickly by touching the function buttons(pads) on the screen, and also to check
    the state of cycle progressing such as the actual temperature, pressure and
    humidity in the chamber etc. conveniently in real time on the screen.

  • 09

    It is effective to sterilize the bulky loads or a large quantity of medical devices and supplies at once by the large sized chamber.

  • 10

    The door is divided to the automatic and manual operating type and user can operate it on the control panel. The safety and sealing function is excellent
    because the door locking unit is composed of the radial lever structure. In case
    the driving unit of automatic operating door is out of order or power failure, user
    is able to operate the door manually by using the door hand wheel without any
    special tool and procedures.

  • 11

    The built-in panel printer prints out the cycle performance record for analyzing the results of

Specifications

Model HG-1000E·P
Style

Single Door 

Overall Dimension W1,385 x D1,692 x H1,894 ㎜
Chamber W680 x D1,266 x H1,200 (㎜)
Vol: 1,033ℓ, Material; STS 316L
Door Operation

Automatic / Manual

Loading Equipment Loading Car : W600 x D1,200 x H1,072 mm
Carriage : W560 x D1,490 x H986 mm
Sterilant 100% Ethylene Oxide(EO), Gas Cartridge type (Gas Weight; 225g x 2 ea)
Water Source Pressure: 1.5~3.0 bar
Power Source AC 230V, 3ph, 3wire, 50/60Hz, 35A
OR AC 380V, 3ph, 4wire, 50/60Hz, 20A
Power Consumption 15kW
Controls Microprocessor
Air Removal Device Vacuum Pump
Display Touch Screen type Color LCD(LED back-light)
Panel Printer Direct Thermal Printing type, Paper Roll; W58 x ∅40㎜ (length: 13 m)
Weight Main body : 1,390kg
Loading Car : 58 kg
Carriage : 52 kg
Environment Conditions ㆍ Altitude : above 2,000 m
ㆍ Temperature Rang : +5℃~+30℃
ㆍ Relative Humidity : max. 80%
ㆍ Power Supply : Fluctuation Range ±10%
ㆍ Installation Category (Overvoltage Category) : Ⅱ, Pollution Degree; 2